Calibration of Results of Water Meter Test Facility
Andrius Bončkus
Gediminas Gediminas
Gediminas Gediminas
Abstract
The results of water meter test facility calibration are presented. More than 30 test facilities are used in Lithuania nowadays. All of them are certificated for water meter of class 2 verification. The results of inter-laboratory comparison of multi-jet water meter calibration at flow rate Q = 5 m3/h are presented. Lithuanian Energy Institute was appointed as reference laboratory for the comparison. Twelve water meter verification and calibration laboratories from Lithuania participated in the ILC. The deviations from reference values were described by the normalized deviation En.
Article in Lithuanian