Research of Relation of Samplers Frequency Characteristics

Tomaš Tankeliun (Lithuania)

Abstract


This paper proposes an algorithm to reduce limitations in band-width measurements of sampling oscilloscopes then using a swept sine-wave measurement method. The traditional swept sine-wave method allow measure only magnitude response. Phase response can be computed only if a sampler is a minimal phase circuit. In this paper alternative bandwidth measurement algorithm using the nose-to-nose method with measurements corrections for the non-idle properties of oscilloscope is de-scribed. Algorithm includes for noise, time base distortions and jitter in measurement signals corrections methods. Proposed algorithm allows to measure phase and magnitude responses when only two similar oscilloscopes and the source of sync pulse are used. Algorithm performs as well as the swept sine-wave method in case when both samplers have the same frequency characteristics.

Article in: Lithuanian

Article published: 2016-06-29

Keyword(s): sampling oscilloscopes; frequency response; minimal phase cir-cuit; Hilbert transform; sampler.

DOI: 10.3846/mla.2016.933